FY342E-IV Corona Charging Static Tester

INTRODUCTION

Using Corona charging to measure the electrostatic propensity of various textile fabrics of all types of composition and construction, also sheet materials such as paper, rubber, plastic, composite sheets, etc.

 

STANDARDS:

ISO 18080-1、GB/T 12703.1、FZ/T 01042

 

CHARACTERISTICS:

1, Large color touch screen display, easy menu mode of operation in English and Chinese.

2, Specially designed high voltage generator circuit ensures continuous and linear adjustment within the range of 0 ~ 10000V. Digital display of high voltage value makes the high voltage adjustment easy and clear.

3, The high voltage generator circuit with fully closed module structure,  switches on and off the high voltage, no more contact ignition, safer with more reliability.

4, The decay time of static voltage is arbitrarily selectable: 1% ~ 99%.

5, Equipped with  humidity sensor,  and dehumidifying air duct, electrostatic half-decay test can be realized in laboratory environment.

6, Both pre-set timed and pre-set voltage method can be used respectively for testing. Digital display of the instantaneous peak value when high voltage discharges, half-decay value (or residual static voltage value) and decay time,  automatic shut down of the high voltage and the motor, easy to operate.

 

TECH. PARAMETERS:

1. Range of electrostatic voltage: 0~10KV

2. Timing range of half-decay: 0 to 99min 99s, error ± 1s

3. Turntable speed: 1400r/min

4. Charging time: 0 ~ 99.99s adjustable (Standard requirement: 30s ± 0.1s)

5. Distance between needle electrode and specimen: 20mm

6. Distance between detective electrode and specimen: 15mm

7. Specimen size: 60mm×80mm×3

8. Humidity sensor test range: 0%RH ~ 100%RH, precision: ±3%

9. Power: 220V, 50HZ, 100W

10. Dimensions: 600×600×500mm (L×W×H)

11. Mass: about 40Kg